Bandung 23,3 °C 1014,5 hPa

Metrology

Kalibrasi SPRT ITS-90

Koefisien fungsi deviasi untuk termometer resistansi platina standar dari titik tetap atau titik komparasi, beserta konversi R <-> T90.

Setup

Fixed points: exact solution from ITS-90 fixed-point readings. Comparison points: least-squares fit to points at arbitrary temperatures.

Sub-ranges to calibrate

Tick one or more ITS-90 sub-ranges. Sub-range 6 is not available in comparison mode.

Measured resistance in the water triple point cell (273.16 K), used in both methods.

Fixed-point method only. Immersion depth l below the liquid surface in the triple point cell; leave empty for no correction.

Fixed point: e-H2 TP (13.8033 K)

Needed for sub-ranges 1, 2.

Optional immersion depth correction.

Fixed point: e-H2 ~17.035 K

Needed for sub-range 1.

Actual T90 of this point; empty = 17.035 K.

Fixed point: e-H2 ~20.27 K

Needed for sub-range 1.

Actual T90 of this point; empty = 20.27 K.

Fixed point: Ne TP (24.5561 K)

Needed for sub-ranges 1, 2.

Optional immersion depth correction.

Fixed point: O2 TP (54.3584 K)

Needed for sub-ranges 1, 2, 3.

Optional immersion depth correction.

Fixed point: Ar TP (83.8058 K)

Needed for sub-ranges 1, 2, 3, 4.

Optional immersion depth correction.

Fixed point: Hg TP (234.3156 K)

Needed for sub-ranges 1, 2, 3, 4, 5.

Optional immersion depth correction.

Fixed point: Ga MP (302.9146 K)

Needed for sub-ranges 5, 11.

Optional; empty = no pressure correction (reference 101325 Pa).

Optional immersion depth correction.

Fixed point: In FP (429.7485 K)

Needed for sub-ranges 9, 10.

Optional; empty = no pressure correction (reference 101325 Pa).

Optional immersion depth correction.

Fixed point: Sn FP (505.078 K)

Needed for sub-ranges 6, 7, 8, 9.

Optional; empty = no pressure correction (reference 101325 Pa).

Optional immersion depth correction.

Fixed point: Zn FP (692.677 K)

Needed for sub-ranges 6, 7, 8.

Optional; empty = no pressure correction (reference 101325 Pa).

Optional immersion depth correction.

Fixed point: Al FP (933.473 K)

Needed for sub-ranges 6, 7.

Optional; empty = no pressure correction (reference 101325 Pa).

Optional immersion depth correction.

Fixed point: Ag FP (1234.93 K)

Needed for sub-range 6.

Optional; empty = no pressure correction (reference 101325 Pa).

Optional immersion depth correction.

Comparison points

Comparison method only. Temperatures in degC, separated by spaces, commas or new lines. Must pair one-to-one with the resistances below.

Comparison method only. Resistances in the same order as the temperatures.

Comparison method only. Widen each sub-range limit by this many kelvin (empty = 0).

Measurements

Optional. Resistances separated by spaces, commas or new lines.

Optional. T90 values in degC separated by spaces, commas or new lines.

Jam tersinkronisasi

Waktu saat ini

.000

UTC

Zona waktu Anda

.000

UTC · Waktu tersinkron yang sama, ditampilkan di zona waktu Anda.

Jam server (NTP), sedang mengukur latensi jaringan Anda…

Latensi pulang-pergi
—
Koreksi jam peramban
—
Perkiraan akurasi
—
Sumber waktu
id.pool.ntp.org (stratum 2)
Offset NTP server
—

Selang acuan

00:00:00.000

Tahan bacaan selang

mulai berhenti

Cara membaca jam ini

Waktu diambil dari server NTP (stratum 1 atau 2 adalah jaringan jam sesium dan GPS). Saat halaman dimuat, browser mengukur perjalanan pulang-pergi ke server ini dan memakai separuhnya sebagai koreksi keterlambatan transmisi. Setelah itu nilainya dimajukan secara lokal; timer JavaScript menyimpang kurang dari satu milidetik per menit pada kondisi normal.

Pakai ini sebagai acuan waktu pengganti frequency counter: rekam layar bersama alat yang diperiksa, lalu baca selangnya dari frame awal dan akhir. Offset NTP yang sama berlaku di kedua ujung sehingga saling menghapus pada selisih; yang tersisa adalah kuantisasi penahanan di atas ditambah jitter jaringan saat pengambilan sampel.