What it does
Calculates the ITS-90 deviation-function coefficients (a, b, c, d) of a standard platinum resistance thermometer (SPRT) for one or more sub-ranges, and converts resistance to temperature (R → T90) and temperature to resistance (T90 → R) with the result.
Two methods:
- Fixed points (exact): the coefficients are solved exactly from readings at ITS-90 fixed points, with optional gas-pressure and immersion-depth corrections.
- Comparison points (least squares): the coefficients are fitted to readings at arbitrary temperatures, using only points inside each sub-range. With exactly as many points as coefficients the fit is exact. The residual of every point is reported in mK.
How to use
- Choose the method and tick the sub-ranges to calibrate.
- Enter R at the water triple point (the cell at 273.16 K).
- Fixed points: fill in the resistance of every fixed point the chosen sub-ranges need (each field says which sub-ranges use it). Pressure (open melting/freezing points only) and immersion depth are optional. The two e-H2 points at about 17.035 K and 20.27 K take an optional actual T90.
- Comparison points: enter the t90 values (°C) and resistances as two lists of the same length. Optionally widen the sub-range limits by some kelvin. Sub-range 6 is not available in this mode.
- Optionally enter resistances (Ω) to convert to T90 and temperatures (°C) to convert to R.
Results: the Rtpw to enter in the instrument (corrected for TPW depth), the W and ΔW at each point, the coefficients per sub-range, and the conversions. The CSV contains all result rows.
- Reference function, T90 < 273.16 K: ln Wr = Σ Aᵢ·[(ln(T90/273.16) + 1.5)/1.5]ⁱ, i = 0…12.
- Reference function, T90 ≥ 273.16 K: Wr = Σ Cᵢ·[(T90 − 754.15)/481]ⁱ, i = 0…9.
- W = R / Rtpw, ΔW = W − Wr(T90).
- Fixed-point temperature shift: ΔT = (dT/dl)·l + (dT/dp)·(p − 101325 Pa), pressure term for open points only. Then W = R/Rtpw − (dWr/dT)·ΔT. Rtpw = R_measured / (1 + (dWr/dT)·ΔT_TPW).
- Deviation functions (W − 1 = x):
- Sub-ranges 1–4: ΔW = a·x + b·x·ln W + Σ cᵢ·(ln W)ᵏ, with 5, 3, 1 or 0 extra terms depending on the sub-range.
- Sub-ranges 5, 8, 9: ΔW = a·x + b·x².
- Sub-ranges 7: ΔW = a·x + b·x² + c·x³. Sub-range 6 adds d·(W − W_Al)² for W > W_Al.
- Sub-ranges 10, 11: ΔW = a·x.
- R → T90: Wr = W − ΔW(W), then T90 is found by inverting Wr (bisection). T90 → R: W = Wr(T90) + ΔW(W), solved by iteration.
FAQ
Which fixed points do I need? Each field lists the sub-ranges that need it. If one is missing, the tool tells you which.
What does "out of range" mean in a conversion? The temperature found lies outside the limits of that sub-range, so that sub-range's calibration does not cover it.
Why is the result different from my certificate? Check that the pressure and depth corrections match those used by the laboratory, and that Rtpw is the value measured in your cell.
What are the limits of the least-squares fit? At least as many points other than W = 1 as coefficients are needed in each sub-range.
Credit and license. Original method and script by Mughni Yumashar (mughnimail@gmail.com). Non-commercial use requires credit to the author; commercial use requires the author's consent.