Bandung 23.3 °C 1014.5 hPa

Metrology

Surface Plate Flatness (Full Grid + Monte Carlo) as per SNSU PK.P-08:2022

Computes the flatness of a surface plate from electronic level gradients measured on a crossing grid (least-square solution), fits the optimal datum plane and evaluates the measurement uncertainty with a Monte Carlo simulation.

Grid

Number of grid segments along the x direction (plate length divided by the step length l, clause 5.1.5). The guide covers 3 to 13 base lengths per side (clause 2.3).

Number of grid segments along the y direction (plate width divided by the step length l, clause 5.1.5).

Distance between two neighbouring grid points, equal to the base length of the electronic level (clause 5.3.1). Example: 0.15 for a 150 mm base.

Measurement

Mean gradient of every grid segment, in measurement order (clause 5.1.6.b). Point (0,0) is the top-left corner, x runs downward, y to the right. First the m*(n+1) segments parallel to x: for each y-line iy=0..n and each ix=0..m-1 the segment (ix,iy)-(ix+1,iy). Then the (m+1)*n segments parallel to y: for each y-line iy=0..n-1 and each ix=0..m the segment (ix,iy)-(ix,iy+1). Total number of values = m*(n+1) + (m+1)*n.

Instrument information

Name or type of the surface plate (report information).

Manufacturer of the surface plate (report information).

Serial number of the surface plate (report information).

Date of the calibration (report information).

Place of the calibration (report information).

Reference standard used for the calibration, e.g. the electronic level and its calibration certificate (traceability statement, report information).

Uncertainty sources

Half-width of the rectangular distribution of the determination of the electronic level base length; systematic and correlated between segments. Example value: 0.005 m.

Repeated gradient readings of one and the same segment. Their sample standard deviation is used as the standard uncertainty of repeatability (random, independent for every segment). At least 2 values.

Half-width of the rectangular distribution of the rounding of the electronic level reading, i.e. half of the display resolution (1 µm/m resolution -> 0.5 µm/m); random and independent for every segment.

Standard deviation of the normal distribution of the correction value of the electronic level; systematic and correlated between segments. Example value: 2 µm/m.

Half-width of the rectangular distribution of the residual error of the applied regression equation; random and independent for every segment. Example value: 0.97 µm/m.

Number of Monte Carlo iterations. The guide uses 200000 as an example; large values on a large grid take longer to compute.

Optional seed, for reproducible Monte Carlo results.

Synchronised clock

Current time

.000

UTC

Your time zone

.000

UTC · The same synchronised instant, shown in your time zone.

Server clock (NTP), measuring your network latency…

Round-trip latency
—
Browser clock correction
—
Estimated accuracy
—
Time source
id.pool.ntp.org (stratum 2)
Server NTP offset
—

Reference interval

00:00:00.000

Hold interval reading

start stop

How to read this clock

The time comes from an NTP server (stratum 1 or 2 is a network of caesium and GPS clocks). On page load the browser measures the round trip to this server and applies half of it as a correction for transmission delay. After that the value is advanced locally, and JavaScript timers drift less than a millisecond per minute under normal conditions.

Use this as the time base instead of a frequency counter: film the screen together with the device under test, then read the interval from the start and stop frames. The same NTP offset applies to both ends, so it cancels in the difference; what remains is the hold quantisation above plus network jitter at the moment of sampling.