What it does
Solves the ITS-90 deviation-function coefficients (a, b, c, d) of a standard platinum resistance thermometer (SPRT) exactly from readings at ITS-90 fixed points. Gas-pressure and immersion-depth corrections are optional. The coefficients can then be used in the Conversion R ↔ T90 tool.
How to use
- Tick the sub-ranges to calibrate.
- Enter R at the water triple point (273.16 K). Optionally enter the sensor depth in the TPW cell.
- Fill in the resistance of each fixed point the chosen sub-ranges need (each field says which sub-ranges use it). If one is missing, the tool tells you which.
- Optional per point: immersion depth (m). For the open metal points (Ga, In, Sn, Zn, Al, Ag) also the gas pressure (Pa); empty = no pressure correction (reference 101325 Pa).
- The two e-H2 points (~17.035 K and ~20.27 K) take an optional actual T90 in K; empty = the nominal value.
| Sub-range |
Range |
Fixed points needed |
| 1 |
13.8033 K – 273.16 K |
e-H2, e-H2 ~17 K, e-H2 ~20 K, Ne, O2, Ar, Hg |
| 2 |
24.5561 K – 273.16 K |
e-H2, Ne, O2, Ar, Hg |
| 3 |
54.3584 K – 273.16 K |
O2, Ar, Hg |
| 4 |
83.8058 K – 273.16 K |
Ar, Hg |
| 5 |
234.3156 K – 302.9146 K |
Hg, Ga |
| 6 |
0 °C – 961.78 °C |
Sn, Zn, Al, Ag |
| 7 |
0 °C – 660.323 °C |
Sn, Zn, Al |
| 8 |
0 °C – 419.527 °C |
Sn, Zn |
| 9 |
0 °C – 231.928 °C |
In, Sn |
| 10 |
0 °C – 156.5985 °C |
In |
| 11 |
0 °C – 29.7646 °C |
Ga |
Results: the Rtpw to enter in the instrument (corrected for TPW depth), the corrections applied in mK, W and ΔW at each point, and the coefficients per sub-range (labelled a4, b4, c7, …; for sub-ranges 1–3 the c terms are c1, c2, …). The CSV holds every result row.
- W = R / Rtpw, ΔW = W − Wr(T90), with the reference function Wr from ITS-90 (Eq. 9a below 273.16 K, Eq. 10a above).
- Temperature shift at a fixed point: ΔT = (dT/dl)·l + (dT/dp)·(p − 101325 Pa); the pressure term applies to open points only.
- Corrected point: W = R/Rtpw − (dWr/dT)·ΔT. Corrected Rtpw = R_measured / (1 + (dWr/dT)·ΔT_TPW).
- Deviation functions, with x = W − 1:
- Sub-range 1: ΔW = a·x + b·x·ln W + c1·(ln W)³ + c2·(ln W)⁴ + c3·(ln W)⁵ + c4·(ln W)⁶ + c5·(ln W)⁷
- Sub-range 2: ΔW = a·x + b·x·ln W + c1·ln W + c2·(ln W)² + c3·(ln W)³
- Sub-range 3: ΔW = a·x + b·x·ln W + c1·(ln W)²
- Sub-range 4: ΔW = a·x + b·x·ln W
- Sub-ranges 5, 8, 9: ΔW = a·x + b·x²
- Sub-range 7: ΔW = a·x + b·x² + c·x³
- Sub-range 6: as sub-range 7, plus d·(W − W_Al)² for W > W_Al; a, b, c come from Sn, Zn, Al and d from Ag
- Sub-ranges 10, 11: ΔW = a·x
- The coefficients solve a linear system with one equation per fixed point (Gaussian elimination).
FAQ
What does "singular system" mean? The fixed-point data cannot give a unique solution. Check for duplicated or wrong values.
Which pressure do I enter? The gas pressure in the cell at the plateau, in pascals. Leave it empty if you do not correct for it.
Why is Rtpw different from what I entered? Because of the TPW depth correction. With no depth entered, it is the same value.
My result differs from the certificate. Check that the pressure and depth corrections match those used by the laboratory.