What it does
Converts resistance to temperature (R → T90) and temperature to resistance (T90 → R) for one sub-range, using deviation-function coefficients you already know, for example from a calibration certificate or from the Fixed Points and Comparison Points tools.
How to use
- Choose the sub-range the coefficients belong to.
- Enter Rtpw, the water triple point resistance used with those coefficients.
- Fill in the coefficients the sub-range needs. If one is missing, the tool tells you which.
- Optional: extend the sub-range limits by some kelvin (for example if the coefficients came from a comparison with extended limits).
- Enter the resistances (Ω) to convert to T90 and/or the temperatures (°C) to convert to R. At least one value is needed.
| Sub-range |
Coefficients to enter |
| 1 |
a, b, c1, c2, c3, c4, c5 |
| 2 |
a, b, c1, c2, c3 |
| 3 |
a, b, c1 |
| 4, 5, 8, 9 |
a, b |
| 7 |
a, b, c |
| 6 |
a, b, c, d and W_Al (W of the Al freezing point) |
| 10, 11 |
a |
Coefficients entered but not used by the chosen sub-range are ignored, with a warning.
Results: the coefficients used, then for each resistance the W, ΔW, t90 (°C and K) and whether it is "in range"; for each temperature the R, W and ΔW. The CSV holds every result row.
- R → T90: W = R / Rtpw; Wr = W − ΔW(W); T90 is found by inverting the ITS-90 reference function Wr (bisection between 13.8033 K and 1234.93 K).
- T90 → R: Wr(T90) is computed, then W = Wr + ΔW(W) is solved by iteration; R = W · Rtpw.
- Deviation functions (x = W − 1):
- Sub-range 1: ΔW = a·x + b·x·ln W + c1·(ln W)³ + c2·(ln W)⁴ + c3·(ln W)⁵ + c4·(ln W)⁶ + c5·(ln W)⁷
- Sub-range 2: ΔW = a·x + b·x·ln W + c1·ln W + c2·(ln W)² + c3·(ln W)³
- Sub-range 3: ΔW = a·x + b·x·ln W + c1·(ln W)²
- Sub-range 4: ΔW = a·x + b·x·ln W
- Sub-ranges 5, 8, 9: ΔW = a·x + b·x²
- Sub-range 7: ΔW = a·x + b·x² + c·x³
- Sub-range 6: as sub-range 7, plus d·(W − W_Al)² for W > W_Al
- Sub-ranges 10, 11: ΔW = a·x
FAQ
What does "out of range" mean? The temperature found lies outside the limits of the sub-range, so its calibration does not cover that value.
What does "no calibrated sub-range covers this result" mean? Same as above, for the only sub-range chosen. Pick a sub-range that covers your temperature, or extend the limits.
A temperature is rejected. t90 must lie within 13.8033 K to 1234.93 K (−259.3467 °C to 961.78 °C).
Which Rtpw do I use? The one the coefficients were calculated with. The Rtpw to enter in the instrument from the calibration tools is the right choice.
Can I convert with several sub-ranges at once? Not in one run; run the tool once per sub-range and compare the results.