What it does
Fits the ITS-90 deviation-function coefficients (a, b, c) of an SPRT to readings taken at arbitrary temperatures (for example against a reference thermometer), by ordinary least squares on ΔW = W − Wr. The residual of every point is reported in mK. The coefficients can then be used in the Conversion R ↔ T90 tool.
How to use
- Tick the sub-ranges to calibrate. Sub-range 6 is not available here, because its d term needs the Al fixed-point reading itself.
- Enter R at the water triple point (273.16 K / 0.010 °C).
- Enter the t90 values (°C) and the resistances (Ω) as two lists of equal length, in the same order. Separate values by spaces, commas or new lines.
- Optional: extend the sub-range limits by some kelvin, so points slightly outside a sub-range are still used.
Only points inside a sub-range (plus the extension) are used for that sub-range. Points at W = 1 (the water triple point itself) carry no information and are shown but not counted in the fit. Each sub-range needs at least as many other points as it has coefficients:
| Sub-range |
1 |
2 |
3 |
4 |
5 |
7 |
8 |
9 |
10 |
11 |
| Coefficients |
7 |
5 |
3 |
2 |
2 |
3 |
2 |
2 |
1 |
1 |
Results: Rtpw to enter in the instrument, the fit type ("exact solution" when the point count equals the coefficient count, otherwise "least-squares fit"), W, ΔW and residual (mK) for each point used, the coefficients, and the points not used. The CSV holds every result row.
- W = R / Rtpw, ΔW = W − Wr(T90); Wr is the ITS-90 reference function.
- Least squares: the coefficients minimise Σ (ΔW − ΔW_fit)² over the points used (normal equations solved by Gaussian elimination).
- Deviation functions (x = W − 1): the same as in the fixed-point tool, e.g. sub-range 4: ΔW = a·x + b·x·ln W; sub-range 7: ΔW = a·x + b·x² + c·x³.
- Residual (mK) = (ΔW − ΔW_fit) / (dWr/dT) × 1000.
- Rtpw is used as entered, with no depth correction.
FAQ
Why "needs at least N points other than W = 1"? A sub-range cannot be solved with fewer informative points than coefficients. Add points inside the sub-range, or extend its limits.
Why are some points "not used"? They lie outside the limits of that sub-range. This is normal when you calibrate several sub-ranges from one data set.
Are large residuals a problem? They show how well the fit reproduces each point. Large values suggest a bad reading or a poor choice of sub-range.
A temperature is rejected. t90 must lie within 13.8033 K to 1234.93 K (−259.3467 °C to 961.78 °C).